EMS-Trade
Cooling
Loop consumables

Document ECT-01 · REV C · ISSUED 2026-09 Six lines · 48 sheets

Note · NB-03 · cleanliness

ISO 4406 in practice: what 17/15/12 actually removes

The target is three code numbers read at one sampling, and it says nothing on its own about where the sample came from. Draw it from the tank and the loop reads clean while the element is still loading. Draw it on the return, upstream of the charge, and the same loop reads the number it is.

Target ISO 4406 17/15/12 Absolute ladder 0.05 µm to 100 µm Fine stage 1 µm Change point 0.5 – 0.7 bar

Request this note as a sheet Sampling, rating ladder, element schedule
A pleated filter element cut through the middle on a bench, the pleat pack packed with dark solids on the outer face and still open towards the bore
FIG 1 ·

An element cut open is the only honest record of what a loop was carrying. The count is a photograph of one moment; the element is the roll of film.

01

The target is a snapshot, not an average

Claim · and the one habit that makes the number mean something

A cleanliness figure is a statement about fluid at one place at one moment. It is not a property of the loop and it does not average out. Two samples drawn an hour apart from the same loop, one at the tank and one on the return, can honestly differ by a wide margin.

The useful discipline is not to chase a lower number. It is to draw every sample from the same tapping, at a comparable point in the cycle, and write both down beside the figure. A number without its tapping is a claim that cannot be repeated.

What the target is for: ISO 4406 17/15/12 before the first charge. After that it is the reference the loop is held against, and the figure compared is the one drawn at the same tapping.

A sample tapping on a stainless return line with a small valve on a boss and a capped bottle held under it, the pipe insulation cut back around the boss
FIG 2 ·

The tapping is part of the instrument. Move it and every earlier reading becomes incomparable.

A clean sample bottle with its cap off standing on a bench beside a second bottle already filled, both unlabelled and both from the same line
FIG 3 ·

Two bottles from two tappings describe two different fluids. Both are true, and only one is comparable.

A return line running back along a plant room wall with a filter housing in the run, the pipe insulated and the housing lid bolted down
FIG 4 ·

On the return, upstream of the charge connection, the sample has been through the loop rather than sitting in it.

A tank lid standing open on a reservoir with the fluid inside still, and a sample bottle resting on the rim beside it
FIG 5 ·

A still tank lets solids settle. A sample from the top of it describes the fluid that has already been cleaned.

02

What sets the number, and what only reports it

Mechanism · the rating removes, the sampling point reports

The rating removes

An absolute rating states the largest particle that passes. Nothing downstream can be cleaner than the finest stage the fluid has been through, which is why the figure is chosen once.

The pressure reports

Differential pressure across the housing rises as the element loads. Between 0.5 bar and 0.7 bar it says the element is done, and a count that keeps improving past that point is fluid bypassing rather than cleaning.

The water reports too

Conductivity and pH sit on the same schedule as the particle count: a fluid that has taken on top-up water reads differently on all three, and the particle number alone will not say which moved.

How one number is built

A code carries three numbers because the count is taken in three size bands at one sampling. One band can improve while another holds, so a single figure would hide the change that matters. Read as a set, they say whether the loop is getting cleaner or moving what it carries between bands.

3 Bands, one sampling Read together, never separately
1 µm Usual fine stage Absolute rating
0.5 – 0.7 bar Change point Across the housing
Two filter elements side by side on a bench, one new with white pleats standing sharp, one loaded with the pleats packed together and greyed
FIG 6 ·

New and loaded, same element. The change in appearance is what the pressure across the housing is measuring.

A differential pressure gauge piped across a filter housing on a plant room skid, two impulse tubes running from the gauge down to either side of the housing
FIG 7 ·

Two impulse tubes, one gauge, and a reading that is only meaningful against the clean zero the element started at.

A stainless filter housing with its bolted lid lifted clear, the element bore visible inside and the lid gasket resting on the rim
FIG 8 ·

Every element change is an opportunity to open the loop. The lid seal is the part that pays for it.

03

From 100 µm down to 0.05 µm

Schedule · the eight absolute steps, and where each one belongs

L1 ·

100 µm

The coarse end. Holds the debris a first fill brings with it: swarf, packaging fibre, seal fragments from assembly.

First pass on a new build

L2 ·

10 µm

The bag stage on a loop with known debris in it. Cheap to change, and it takes the load off the fine stage behind.

Pre-filter on a dirty fill

L3 ·

5 µm

Where a loop starts being clean rather than merely less dirty. Useful as the first stage while a flush is running.

Mid flush

L4 ·

1 µm

The usual fine stage and the figure a loop is quoted on. It also loads fastest, so it is the step the pressure gauge is watching.

In service

L5 ·

0.45 µm

Beyond the loop. Used on a polishing side stream or a sample line, where the flow is small enough not to cost much pressure.

Side stream

L6 ·

0.2 µm

Polishing on a small branch only. On a full loop the pressure drop decides when it comes out.

Small branch

L7 ·

0.1 µm

A bench figure for a laboratory loop, not a facility one. Worth knowing so a specification asking for it can be questioned.

Laboratory scale

L8 ·

0.05 µm

The bottom step. Asking for it on a facility loop moves the question from filtration to hydraulics, and the answer changes with it.

The bottom step

Eight pleated elements of the same length standing in a row on a bench, the pleat spacing visibly tighter from one end of the row to the other
FIG 9 ·

The ladder as stock. Two steps of it usually go into a loop; the other six exist so the two can be chosen rather than inherited.

04

The figures that decide it

Schedule · value, unit and the condition it holds under

Cleanliness figures, by what they decide
Item Figure Value Unit Condition it holds under
01 Loop targetBefore the first charge ISO 4406 17/15/12 Three code numbers, read as a set at one sampling. A higher class in any one of them is a different target.
02 Fine stageElement 1 µm Absolute rating. The step that holds the target and the step that loads fastest.
03 Coarse end of the ladderElement 100 µm Takes the load of a first fill so the fine stage is not spent on debris a coarse element could have held.
04 Element change pointHousing 0.5 – 0.7 bar Differential across the housing, measured from the clean zero the element was run to before it shipped.
05 Housing design pressureShell ≥ 1.0 MPa On the shell, above the working pressure of the loop it sits in.
06 Element lengths heldCartridge 20 / 40 in Standard lengths, with a short element on a small branch where full flow would cost too much pressure.
07 ConductivityWater µS/cm range Read on the same sample as the particle count, never as a separate visit.
08 pHWater pH scale Moves when the fluid ages rather than when it is diluted. The pair is what tells the two apart.
09 System certificateAcross the lines ISO 9001 certified Held across the supply network rather than by any one line. Verified before a line is scheduled.
10 Material certificateMetal parts EN 10204 3.1 For housings and fittings, traceable to the heat the material came from.

Ten figures. The two read together every time are the particle count and the differential pressure: a count without its pressure cannot tell a clean loop from a bypassed one.

A ruled record sheet on a bench with a sample bottle standing on it and a pen laid across the top edge, the sheet ruled into a column of rows
FIG 10 ·

The figure that gets repeated is the one written beside its tapping, its pressure and its date.

05

What to do on the next flush

Practice · five steps, all of them on the flushing contractor's side of the line

STEP ONE Fix the tapping before the first sample On the return, upstream of the charge
STEP TWO Two stages, coarse first The fine element is not a debris bin
STEP THREE Run to the clean zero and write it down Then read pressure as a rise
STEP FOUR Sample water and particles together One visit, one record
STEP FIVE Cut one element open And keep the photograph
A flushing cart connected to a manifold by two hoses on a plant room floor, the cart frame standing on castors with its hoses coiled at the connection point
FIG 11 ·

The flush is the only time the whole loop passes through one element at a controlled flow.

Flush at the flow the loop will run at

A flush run slowly cleans the pipe it passes through and misses the branch it never enters. The target is the loop's, not the cart's.

Write the record at the time, not afterwards

Three code numbers, the differential pressure and the tapping point, on one line. A record rebuilt from memory the following week is not a record.

Residue control
Flushing fluid left in the loop changes the conductivity of the charge that follows it. The target is a dry, drained loop with a sample to show it.
Who runs it
The flush and the first charge are executed by the contractor commissioning the loop, against a written target. This note is the target and the sampling discipline, not the procedure.
A sample bottle being filled at a tapping on a return line, the bottle held low under the valve with the fluid running down the inside wall
FIG 12 ·

Draw it the same way every time. The method is part of the figure, and so is the person drawing it.

A loaded element opened along its length on a bench with a cutting tool beside it and the two halves laid so the pleat pack faces up
FIG 13 ·

Cutting an element open costs one element and answers the question the count raises: what was it, and where did it come from.

06

The instruments this note is written against

Supplier parts · the two that make the figure repeatable

A pressure transmitter on a dark ground, a white housing with a blue screwed cap on the left and a stainless threaded process connection below
SP 1 ·

A differential transmitter across the housing is what turns the element from a calendar item into a measurement. It is read as a rise from the clean zero, not as an absolute.

<A mobile filter trolley on castors carrying two grey cylindrical filter bodies with flanged lids, a blue handle on each valve, a pressure gauge and a yellow pump
SP 2 ·

Two bodies on one cart is the two-stage arrangement as hardware: the coarse element in the first body takes the load, the fine one in the second holds the figure.

Both are supplier parts from the six lines on this site. The water side is scheduled on fluid quality monitoring, and the elements themselves on filter cartridges.

07

What this note does not cover

Boundary · three things a cleanliness target will not tell you

Not a source
A count says how much is in the fluid and how it divides across three bands. It does not say what the solids are, or whether they came from a hose, a seal or the water used to top up.
Not a cleaning method
Holding a figure is a matter of element rating, flow and change point. Which of the three to move is a decision about the loop, not the fluid, and it is made against the pressure reading.
Not a laboratory method
The code is a counting standard, not an instrument. Any figure here is comparable only with one taken the same way, on the same tapping, by the same method.
A run of insulated pipe along a plant room wall with three tapping bosses at intervals, each with its valve capped and its own pipe marking
FIG 14 ·

Three tappings on one run, three different figures, and all three defensible. Which one is the loop's figure depends on the question being asked.

DOCUMENT REQUEST

Request this note as a sheet

The sampling discipline, the eight-step rating ladder, the ten-figure schedule and the element lengths held across the filtration line.

If the form cannot send, this address takes the request instead: sales@ems-cool.com
The only promise on this page

One business day, with the file attached.

Nothing else on this sheet is promised. The target above is the one this site is indexed on; a stricter one is a different schedule and is quoted against a request.

Three pleated elements of different lengths lying side by side on a bench with the end caps facing the camera and a steel rule laid along the row
PL 1 ·

Length, rating, material and end cap. Four answers and the element is a stock item.

A clipboard with a ruled record sheet on a bench beside two capped sample bottles and a pen, the sheet filled out only in its header row
PL 2 ·

The header row is the part that makes the figures comparable: tapping, date, pressure, then the three numbers.